81
Reference Manual
00809-0100-4811, Rev DA
Section 4: Basic Configuration/Start-Up
January 2015
Basic Configuration/Start-Up
4.6.9 Setup - Probe
The Probe tab contains information on probe type and length. The Rosemount 3300 Series
Transmitter makes some initial calibrations automatically, based on the chosen probe type. The
following probe types are available:
Rigid twin
Flexible twin
Coaxial
Rigid single 0.3 in. (8 mm)
Flexible single
Coaxial HTHP
Coaxial HP/C
Rigid single HTHP 0.3 in. (8 mm)
Flexible single HTHP
Rigid single PTFE
Flexible single PTFE
Rigid single HP/C 0.3 in. (8 mm)
Flexible single HP/C 0.3 in. (8 mm)
Flexible single HP
Rigid single 0.5 in. (13 mm)
User defined
Note
Flexible and rigid probes require different radar electronics and cannot be used with the same
transmitter head.
The probe length is the distance from the upper reference point to the end of the probe, see
Figure 4-1 on page 62. If the probe is anchored to a weight, do not include the height of the
weight.