116
Reference Manual
00809-0100-4811, Rev DA
Appendix A: Reference Data
January 2015
Reference Data
A.1.2 420 mA HART
®
(output option code H)
See Ordering Information in Table A.6 on page 152.
A.1.3 Output
Two-wire, 4–20 mA. Digital process variable is superimposed on 4–20 mA signal, and available to any host
that conforms to the HART protocol (HART rev. 5). The HART signal can be used in a multidrop mode.
Figure A-1. Rosemount 3300 Series Wiring
HART Tri-Loop
By sending the digital HART signal to the optional HART Tri-Loop, it is possible to have up to
three additional 4–20 mA analog signals. See the Rosemount 333 HART Tri-loop Product Data
Sheet (document number 00813-0100-4754) for additional information.
Figure A-2. HART Tri-Loop
A
B
C
D
I
F
E
H
G
A Rosemount 3300 Series Transmitter
B Rosemount 751 Field Signal Indicator
C 4-20 mA/HART
D Field Communicator
E 3 x 4-20 mA
F Rosemount 333 HART Tri-Loop
GHART modem
H Control system
I Radar Configuration Tools or AMS
®
Device Manager
Vista de página 125
1 2 ... 121 122 123 124 125 126 127 128 129 130 131 ... 223 224

Comentários a estes Manuais

Sem comentários