35
Reference Manual
00809-0100-4811, Rev DA
Section 3: Installation
January 2015
Installation
Coaxial
To cut a coaxial probe, do the following:
Note
The HTHP coaxial probe must not be
cut in field.
1. Insert the centering piece.
(The centering piece is delivered
from factory and should be used
to prevent the spacers centering
the rod from coming loose).
2. Cut the tube to the desired length.
3. Move the centering piece.
4. Cut the rod inside the tube. Make
sure the rod is fixed with the
centering piece while cutting.
Pipes longer than
49 in. (1250 mm) can be
shortened by as much as
23.6 in. (600 mm).
Pipes shorter than
49 in. (1250 mm) can be cut to
a minimum length of
15.7 in. (400 mm).
5. Update the transmitter
configuration to the new probe
length, see “Probe length” on
page 70.
Centering piece
Maximum shortening
23.6 in. (600 mm)
L > 49 in.
(1250 mm)
Minimum probe length
25.4 in. (645 mm)
Minimum probe length
15.7 in. (400 mm)
L
49 in.
(1250 mm)
Vista de página 44