Reference Manual
00809-0100-4811, Rev DA
Appendix C: Rosemount 3300 Series with HART
®
to Modbus
®
Converter
January 2015
Rosemount 3300 Series with HART
®
to Modbus
®
Converter
4. Search for and connect to device.
a. Select Device > Search to open the Search Device window.
If the HMC is configured for Modbus ASCII or Levelmaster communication, cycle
the power to the transmitter and then open the Search Device window. (The HMC
will then communicate using Modbus RTU for 20 seconds and under that time it
is possible to connect with RRM.)
b. In the Protocol(s) list, make sure Modbus is selected.
c. Select Scan Address Range, and type a start and end address for HMC Modbus.
The default HMC Modbus address is 246.
d. Select Start Scan.
e. Select OK to connect when the device is found.
5. Make sure the HMC Modbus address is set to 246 (the default address).
a. Select Setup > General.
b. In the Device Address box, type 246 and select Store.
6. Enter service mode.
a. Select Service > Enter Service Mode.
b. Type the password “admin”.
c. Select Submit.