97
Reference Manual
00809-0100-4811, Rev DA
Section 6: Service and Troubleshooting
January 2015
Service and Troubleshooting
The RCT allows you to view a waveform plot to analyze the measurement signal. The plot shows
the signal and the thresholds used for the different amplitude peaks. By adjusting amplitude
threshold T3, it is possible to detect even weak interface signals.
Guidelines for amplitude threshold settings:
The amplitude threshold T3 should be approximately 50 % of the interface signal
amplitude.
Threshold T3 should not be less than 3.
If possible, T3 should be higher than T2.
You can use the RCT software or a Field Communicator to change the amplitude thresholds. For
the Field Communicator, use the HART command [1, 4, 5, 3]. See also Amplitude threshold
settings” on page 104.
RCT allows you to view a plot of the measurement signal along with the current thresholds:
1. From the View menu, select the Plotting option, or double-click the Plotting icon in
the Tools section of the RCT project bar.
2. Select the Read button .
3. To adjust the amplitude thresholds, open the Tools section in the RCT project bar and
select Device Commands > Details > Set Nominal Thresholds.
Figure 6-4. Waveform Plot Indicating that the Amplitude Threshold for the Interface Peak
is Too High
Figure 6-4 on page 97 illustrates a situation where amplitude threshold T3 is too high. The signal
amplitude peak at the interface between the upper and lower products is not detected in this
case. By adjusting amplitude threshold T3, the peak at the interface between the upper and
lower products is detected as illustrated in Figure 6-5:
The amplitude threshold is above
the measurement signal peak